AS 4862:2000 (R2016)
$27.30
Lead and lead alloys – Determination of impurities and alloying elements – Atomic emission spectrometric method
Published By | Publication Date | Number of Pages |
AS | 2000-09-04 | 17 |
Sets out a method for the determination of impurities and alloying elements in lead and lead alloys by arc/spark atomic emission spectrometry.
Scope
This Standard specifies a method for the determination of impurities and alloying elements
in lead and lead alloys by arc/spark atomic emission spectrometry. The method is applicable
to the following elements in the concentration ranges shown below:
Element Concentration range, %
Aluminium 0.0001 – 0.05
Antimony 0.0002 – 14.0
Arsenic 0.0003 – 3.0
Bismuth 0.0001 – 1.5
Cadmium 0.0001 – 0.2
Calcium 0.0001 – 1.2
Cobalt 0.0001 – 0.02
Copper 0.0001 – 0.6
Gold 0.0003 – 0.1
Indium 0.0001 – 0.1
Iron 0.0001 – 0.02
Magnesium 0.0001 – 0.5
Manganese 0.0001 – 0.02
Nickel 0.0001 – 0.02
Selenium 0.0005 – 0.06
Silver 0.0001 – 1.0
Sodium 0.0001 – 0.01
Strontium 0.0001 – 0.1
Sulphur 0.0005 – 0.02
Tellurium 0.0002 – 0.3
Thallium 0.0001 – 0.5
Tin 0.0001 – 50
Zinc 0.0001 – 0.1